Flange Resistance Thermometer TP 19 Hengesbach

 The TP19 resistance thermometers with/without transmitter are powerful, rugged devices for process measurements, the food, pharmaceutical and chemical industry, etc., completely made of stainless steel in a sealed design for hygienic applications.

The standard version of the temperature sensor is designed according to Accuracy Class A. Accuracy classes B 1/2, 1/3, 1/10 DIN are optionally available. The measuring inserts are single or double PT 100s, which can be exchanged.







The utilisation of the measuring inserts must be adapted to the requirements with regard to temperature, length, flexibility, vibration resistance and measuring accuracy. Special devices are available for control processes, e.g. tapered sensor tip for particularly short half-value times.

The design according to IP 67 with exchangeable test insert provides access to the measuring point at any time and offers a reliable guarantee for tightness and long operating times, even under extreme operating conditions, as well as saving resources. Programmable 4-20 mA, HART or Profibus PA transmitters can be installed for transmission of the measuring values. The QUICKTEMP series is particularly recommended for measuring points that require calibration.

Features:

  • "fr" housing in ip 67
  • single or double pt100
  • optionally with measuring transmitter
  • connection "varivent" flange, d = 25 or 68 mm
  • fully sealed
  • short addressing times and tapered sensor tip
  • optional as quicktemp versions for measuring points required for quality assurance

Specification:

Design configuration

   
TypeField housing (Cr Ni St)  
    
Temperature sensorPT 100 according to EN 60751, exchangeable test insert (standard)  
 Insertion tube 1.4571  
    
Test insert, electrical1 x PT 100 in a 2-, 3- or 4-wire circuit, standard: 1 x 3-wire circuit, Class A,  
 2 x PT 100 in 2- or 3-wire circuit  
    
Electrical connectionM 16x1.5 cable connection, MS-nickel-plated,  
 Optional:M12 x 1 round plug, 4-pin 
Accuracy- Standard: Tolerance Class A according to EN 60751,  
 - Optional: Tolerance categories B 1/10, 1/5, 1/3, 1/2  
Response timeT50 (measured in water): 6 s (design-dependent, d = 6 mm),  
 < 6 s with tapered sensor tip, on request)  
    
Temperature of measuring materialtmax at the sensor = -20... +200°C  
Casing materialStainless steel 1.4301 (standard field housing)  
    
Sensor materialStainless steel 1.4571, optional 316L = 1.4435 or 1.4404,  
 Optional: High surface quality  
    
Sensor diameterStandard: Continuously smooth, d = 6 mm,  
 Optional:Tapered sensor tip d = 4 mm(exchangeable test insert)
  Tapered sensor tip d = 3 mm or 2 mm (test insert not exchangeable) 
    
Sensor lengthStandard EL: 35 mm, VARIVENT d = 68 mm,  
 Deviating length on request  
Collared tubeStandard: Neck pipe length = 50 mm, d = 12 mm  
 for extending the distance, depending on the operating temperature,  
 optionally in other lengths  
Type of protectionEN 60529, IP 67 with cable connection, optionally with cable output  
Permitted pressureDN = 10 bar, depending on design and temperature  
CE conformityEMC guidelines are met, CE symbol  

Process connections

   
-- VARIVENT flange d = 68 mm, DN 40/50 for pipe DN 40 – DN 125   
-- Optional: - VARIVENT DN 25, d = 50 mm   
 - VARIVENT DN 10/15, d = 31 mmMaterial No. 1.4404, optional 1.4435 
    

Accessories

   
Parts for assemblyAccessories for VARIVENT installation systems on request  
    

Options

   
Electrical connection- Ceramic connection base  
 - Flexible connection wires  
 - Sheathed cable version (vibration-resistant version)  
 - Transmitter TE 42, programmable, 4...20 mA, 2-wire switch (standard)  
 - Transmitter TE 41, programmable, galvanically separated  
 - Transmitter TE 52, HART  
 - Transmitter TE 82, Profibus PA  
 - Configuration set including adapter and software for TE41 / TE42  
Process connections- Tapered sensor tip D = 3 mm or 4 mm  
  - Exchangeable test insert, d = 3 mm 
  - Neck pipe L = 75 mm, d = 12 mm (please specify other length) 
  - VARIVENT flange DN 25, 316L, fully welded 
CalibrationFactory calibration, calibration certificate (3-point / 5-point) with DKD standard  
CertificatesMaterial certificate according to EN 10204, EHEDG, FDA  

Download Data Sheet:Flange Resistance Thermometer TP 19

Ref: Hengesbach.com