Bamo Level Leak Localizer and Detector BAMOLEAK

Level Leak Localizer and Detector BAMOLEAK

Bamo Measures Sas has been founded in 1978. Its capital is now 400 000 Euros. Since the beginning, the company has undergone a rapid development centering on industrial pH-meters and conductivity meters. Bamo is now well-known as a leader in the French market.

Since 1990, new developments in the Level Measure have given a new spring in the export. To its own product range, the company holds numerous agencies of foreign companies in Europe and all over the world. So we are offering a complete line with always the best possible service.

Bamo is a company that comes from France. This company selling some product. The product of Bamo such as pH and ORP; Chlorine, disinfection; Data processors; Conductivity; Water hardness; Turbidity; Oxygen; Level; Temperature; Flow; Pressure; Plastic valves; Metal valves.

Leak Localizer and Detector BAMOLEAK

BAMOLEAK allows securing work areas where leaks are prejudicial for equipment, by detecting and precisely locating leaks of conductive liquids. The sensor (cable) is laid along the monitored area or pipeline, and, connected to the evaluation unit. Leakage is detected by the converter, which determines, records, and reports information via the analog output, relays, and the RS485 Modbus interface. Two complementary detectors can be connected to the converter such as MAXITOP LWC (Detection in the double-walled tank) or LWC-B.

Features

  • Leak localization
  • Length up to 3,000 meters (12 areas)
  • Automatic recognition of loop parameters
  • Event recording (date / time)
  • Outputs: 4-20 mA, MODBUS RTU, relays

Application

  • Building automation
  • Data centers
  • Semiconductor Industry
  • Clean rooms
  • Archives storage / Libraries / Museums
  • Heating and air conditioning systems
  • Switching centers, computers room

Another Article:

Specification

Power supplyOr 100 … 240 V AC - 50/60Hz 
 Or 10 ... 30 V DC; 12 ... 24 V AC 
Consumption2…5W 
Contact outputs5 contacts, adjustable, potential-free 
 To set as, or N.O. or N.C. contact 
Switching power250 V AC, 2 A / 30 V DC, 1 A
Maximum cable length3,000 m 
Housing46 x 100 x 127 mm 
 Rail mounting (DIN EN 50 022; 35x7.5 mm) 
ProtectionIP40 
Ambient temperature limits-10.. +45 °C
Sensor (cable) temperature limits-50.. +150°C
Connectors temperature limits-20.. +60 °C
Wall-mount cabinet (Option)175 x 200 x 155 mm 
Screw terminalsScrew connectors, cable diam. Max. 1.5 mm² 
Detection areas12 areas; Setting of name/ TAG to each one 
Accuracy of leak localization< 1 %; ±1 m 
Measuring loop1 input for the 4 poles sensor (cable) 
 2 additional inputs for 2 one-point detectors (e.g. MAXITOP LWC or LWC B) 
Power supply to the detection loopsLocalization with sensor (cable): Max. 20 V AC/DC; Max. 5 mA 
 One-point detectors: Max. 30 mA (each) 
Analogue output 4-20 mAActive signal output (with its own loop power) 
 4 ... 5 mA: No alarm 
 5 ... 20 mA: Localization of the detected leak point = [ cable length x (current value - 5mA) ] / 15mA 
 21 mA: Fault, error message 
Adjustable timerWhen detection occurs: 0.5 to ... 5 minutes delay before signaling 
Real time clockintegrated clock with battery, automatic summer time switch can be activated 
RecordsAutomatic record of all events, max. 48 inputs, events can be acknowledged and deleted, individually 
LanguageTo choose between English, German, French, Spanish, Portuguese, Polish 
Key protection of settings4 digits; Can be deactivated 
Signaling interfaceDisplay: 
 2 line-text LCD 
 Status of contact outputs 
 5 LED (multicolor) 
Digital communicationRS 485 / MODBUS 
SélecteurThe rotary switch on the front panel 

Download Data Sheet: Level Leak Localizer and Detector BAMOLEAK

Ref: bamo.eu