Eltra Analyzer ELEMENTRAC ONH-p

Eltra Analyzer ELEMENTRAC ONH-p


Eltra  Analyzer ELEMENTRAC ONH-p | Oxygen / Nitrogen / Hydrogen

The ELEMENTRAC ONH-p determines oxygen, nitrogen, and hydrogen in inorganic samples by inert gas fusion in an impulse furnace with temperatures in excess of 3.000 °C. The ELEMENTRAC ONH-p guarantees precise and fast sample analysis. The analyzer covers a wide range of applications such as metal, ceramics, and other inorganic materials.

The ELEMENTRAC ONH-p can be supplied with up to two infrared cells with different path lengths, accommodating both high and low-level oxygen analyses. Nitrogen and hydrogen concentrations are determined in the ELEMENTRAC ONHp by a robust and sensitive thermal conductivity cell.

Advantages

  • Simultaneous oxygen/nitrogen or oxygen/hydrogen determination with inert gas fusion technique
  • NEW: closed gas management and optimized gas circulation for sensitive ONH determination
  • NEW: use of cost-efficient argon as carrier gas possible
  • NEW: powerful catalyst furnace for precise oxygen measurement
  • NEW: gas flow system with electronic gas flow control and new leakage test
  • NEW: water-cooled sample port system for effective removal of atmospheric gases
  • Flexible configurations and measuring ranges for O, N, and H
  • High sensitivity IR and TC cells with low detection limits
  • Short analysis time
  • Powerful 8,5 kW* impulse furnace for temperatures in excess of 3,000 °C
  • Economic analysis of grains without capsules
  • NEW: chemicals and tubes are hidden behind a door (removable)
  • NEW: powerful software (supporting data and application export, comment fields, and many more)
  • Single and multi-point calibration (linear regression)
  • NEW: cooling via tap water or heat exchanger or chiller
  • New design allows operation in production control and laboratory

Application

Alloys, Cast Iron, Ceramics, Copper, Refractory Metals, Steel, etc.

Another Article:

Specification

Measured elementshydrogen, nitrogen, oxygen
Samplesinorganic
Furnace alignmentvertical
Sample carriergraphite crucibles
Field of applicationceramics, engineering/electronics,
 steel/metallurgy
 electrode impulse furnace (max. 8,5
 KW*), temperatures in excess of
 3,000 °C
Detection methodsolid state infrared absorption for
 oxygen thermal conductivity for
 nitrogen and hydrogen
Typical analysis time120 - 180 s
Chemicals requiredcopper oxide, magnesium
 perchlorate, Schuetze reagents,
 sodium hydroxide
Gas requiredcompressed air, helium 99.995 %
 pure, nitrogen 99.995% pure, argon
 99.995% pure (if required), all gases
 with (2 - 4 bar / 30 - 60 psi)
Power requirements3~ 400 V, 50/60 Hz, max. 8,500 W
Dimensions (W x H x D)57 x 77 x 63 cm
Weight~ 161 kg
Required equipmentbalance (resolution 0.0001g),
 monitor, PC
Optional accessoriescarrier gas purification, external
 chiller, the gas calibration unit
 * limited to 6.8 kW in application
 settings

Ref: eltra.com

Contact us : sales@wmablog.com or visit our E-Commerce

Download Data Sheet: Oxygen / Nitrogen / Hydrogen Analyzer ELEMENTRAC ONH-p