Hach TU5 Series Laboratory Turbidimeters

 Hach is committed to understanding your overall measurement process to provide you and your lab with the right solution for drinking water and pure water applications.

Hach TU5 Series Laboratory Turbidimeters

How do you currently verify your current process instrumentation measurements? Are you able to match your online and lab turbidity measurements?

You could consider spending valuable time and resources developing a more optimized process on your current instrumentation or you could lean on the experts at Hach for an improved optimized technology: the TU5200 benchtop turbidimeter.

The TU5200 is designed to deliver a revolutionary 360° x 90° Detection Technology, engineered to see more of your sample than any other turbidimeter. This unique technology is used in both the TU5 online and TU5 benchtop turbidimeters, allowing you for the first time to remove the uncertainty of which measurement to trust. The 360° x 90° Detection delivers the best low-level precision and sensitivity while minimizing variability from test to test.

Do you spend a lot of time indexing your sample cell and coating the surface area with silicone oil to get the most accurate reading and account for imperfections found on the glass?

The TU5 Series dramatically reduces the time needed to get a turbidity measurement you can rely on, with 98% less online sample surface area to clean, sealed vials for calibration, and the elimination of the need for indexing and silicone oil in the lab. Not to mention, a smaller online sample volume means you will detect events almost immediately.

Specifciations

TU5200

  • Light Source : Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
  • Range :
    EPA
    0 - 700 NTU / FNU / TE/F / FTU
    0 - 100 mg/L
    0 - 175 EBC
    ISO
    0 - 1000 NTU / FNU / TE/F / FTU
    0 - 100 mg/L
    0 - 250 EBC
  • Accuracy : ±2 % plus 0.01 NTU from 0 - 40 NTU; ±10 % of reading from 40 - 1000 NTU based on Formazin primary standard (at 25 °C)
  • Resolution : 0.0001 NTU / FNU / TE/F / FTU / EBC / mg/L
  • Repeatability : <40 NTU: Better than 1% of reading or ±0.002 NTU on Formazin at 25 °C, whichever is greater; >40 NTU: Better than 3.5% of reading on Formazin at 25 °C
  • Stray Light : <10 mNTU
  • Units : NTU, FNU, TE/F, FTU, EBC; mg/L if calibrated with Degrees calibration curve
  • Operating Temperature Range : 10 - 40 °C (50 - 104 °F)
  • Operating Humidity : 80% at 30 °C (non condensing)
  • Sample Temperature : 4 - 70 °C (39 - 158 °F)
  • Storage Conditions : -30 - 60 °C (-22 - 140 °F)
  • Power Requirements (Voltage) : 100 - 240 VAC
  • Power Requirements (Hz) : 50/60 Hz
  • Certifications : CE compliant; US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version; Complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50); Australian ACMA Marking
  • Dimensions (H x W x D) : 195 mm x 409 mm x 278 mm
  • Weight : 2.4 kg (5.29 lbs.)
  • Warranty : 1 year

TU5300sc / TU5400sc

  • Light Source Class : 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
  • Range :
    EPA
    0 - 700 NTU / FNU / TE/F / FTU
    0 - 100 mg/L
    0 - 175 EBC
    ISO
    0 - 1000 NTU / FNU / TE/F / FTU
    0 - 100 mg/L
    0 - 250 EBC
  • Accuracy : ±2% or 0.01 NTU from 0 - 40 NTU; ±10% of reading from 40 - 1000 NTU based on Formazin primary standard
  • Resolution : 0.0001 NTU / FNU / TE/F / FTU / EBC
  • Repeatability : Better than 1% of reading or ±0.002 NTU (TU5300) or ±0.0006 NTU (TU5400) on Formazin at 25 °C (77 °F), whichever is greater
  • Stray Light : <10 mNTU
  • Units : NTU, FNU, TE/F, FTU, EBC
  • Signal Average Time : TU5300sc: 30 - 90 seconds, TU5400sc: 1 - 90 seconds
  • Response Time : TU5300sc: T90 <45 seconds at 100 mL/min, TU5400sc: T90 <30 seconds at 100 mL/min
  • Sample Temperature : 2 - 60 °C (35 - 140 °F)
  • Sample Pressure : 6 bar (87 psi) maximum, compared to air at sample temperature range from 2 - 40 °C (35.6 - 104 °F)
  • Sample Flow Rate : 100 - 1000 mL/min; optimal flow rate: 200 - 500 mL/min
  • Operating Temperature Range : 0 - 50 °C (32 - 122 °F)
  • Operating Humidity : Relative humidity: 5 - 95% at different temperatures, non-condensing
  • Storage Conditions : -40 - 60 °C (-40 - 140 °F)
  • Enclosure Rating : Electronic compartment IP55; all other functional units IP65 with process head/ACM attached to the TU5300sc/TU5400sc instrument
  • Certifications : CE compliant; US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version; Australian ACMA Marking
  • Dimensions : (H x W x D) 249 mm x 268 mm x 190 mm
  • Weight : 5.95 lbs. (2.7 kg); 11 lbs. (5.0 kg) with all accessories
  • Warranty : 1 year

Reference : hach.com

Berbagi:  