Analytical Industries Oxygen Analyzer with Remote Sensor

 A panel mounted oxygen analyzer with a remote sensor allows flexibility of installation. The GPR-1900 gas analyzer is suitable for trace oxygen measurements as low as 0-1 ppm; the GPR-2900 measures O2 up to 100%. The remote sensor block allows a very easy oxygen sensor replacement for simple and fast maintenance and low cost of ownership. All versions can be locked manually to a single range or auto-range. The user is always aware of the oxygen measurement range used via the range ID output.

Analytical Industries Oxygen Analyzer with Remote Sensor


Highlights
  • Measurement ranges from 0-10ppm up to 0-100% O2
  • Barometric pressure and temperature compensation
  • Accuracy of better than 2% of range
  • Two user configurable alarms
  • Range ID output
  • 12-28 V DC or mains powered options
  • XLT sensor options for CO2 backgrounds

GPR-1900

The favourite instrument for trace measurements as low as 0.1 ppm O2. The sensor housing is made from stainless steel to offer a high measurement integrity. The analyzer has a maximum operational range of 0-1,000 ppm but also allows the user to switch temporarily to 0-25% for calibration with ambient air in the absence of a reference gas. The GPR-1900 also displays real time ambient temperature and pressure.

Options :

  • XLT sensor available.
  • Ranges available: 0-10, 0-100, 0-1,000 ppm & 0-1% O2 (0-25% for calibration only)

GPR-2900

For percentage level measurements up to 100% O2 with the same features as the GPR-1900. The sensor housing is made from Delrin as standard.
Options :

  • Stainless steel sensor housing.
  • XLT sensor available.
  • 0-100% O2 (in place of 25%).
  • Ranges available: 0-1%, 0-5%, 0-10% & 0-25% O2

Features

Features

  • Measurement ranges from 0-10ppm up to 0-100% O2
  • Barometric pressure and temperature compensation
  • Accuracy of better than 2% of range
  • Two user configurable alarms
  • Range ID output
  • 12-28 V DC or mains powered options
  • XLT sensor options for CO2 backgrounds

Applications

  • Nitrogen or oxygen generators
  • Glove boxes
  • Oxygen injection in fish farms
  • Confined space monitoring
  • Sintering and other metal processing
  • Delignification and bleaching in pulp & paper

Specifications

 GPR-1900GPR-2900
Measurement range0-10, 0-100, 0-1000 ppm, 0-1%, 0-25%0-1%, 0-5%, 0-10%, 0-25%;
optional 0-1%, 0-5%, 0-10%, 0-100%
Accuracy< 2% of selected range at constant conditions
Response timeT90 < 10 seconds
Recovery time60 seconds in air to < 10 ppm in < 1 hr on N2 purgeNA
Sensitivity (LDL)0.05 ppm0.005%
Linearity< 0.5% of scale
Sensor modelGPR-11-60-4
XLT-11-24-4 for gases containing > 0.5% CO2
Sensor life 25ÂșC (77°F) and 1 atm24 months in < 1000 ppm O2
6 months in air
GPR-11-60-4 60 months in air
XLT-11-24-4 24 months in air
Calibration interval30 days
Inlet pressure0.34 2 barg (5-30 psig) with atmospheric vent
Flow rate0.5 1.0 l/m (1-2 SCFH)
Gas connections1/8" compression tube fittings
Wetted partsStainless steelDelrin; stainless steel optional
DisplayGraphical LCD 12.7 x 7cm (5 x 2.75”); resolution 0.001 ppm GPR-1900 MS2, 0.01 ppm GPR-1999, 0.001% GPR-2900
EnclosurePainted aluminum 6" x 4" x 4" panel mount
CompensationBarometric pressure and temperature
Signal output4-20 mA or 1-5V
Range ID4-20 mA non-isolated or 1-5V
Option: reassign relay contacts (eliminates standard alarms).
AlarmsTwo user configurable alarms: magnetic coil relays rated 3A at 100 V AC
Operating temperatureGPR sensors: 5°C to 45°C (41°F to 113°F)
XLT sensors: -10°C to 45°C (14°F to 113°F)
Power100-240 V AC or 12-28 V DC

Download Datasheet : Analytical Industries Oxygen Analyzer with Remote Sensor

Reference : https://aii1.com/