Maximize installation and application flexibility with the coplanar platform
■ Improve reliability and performance with integrated DP Flow meters, DP Level solutions, and integral manifolds.
■ Easy installation with all solutions fully assembled, leak-tested, and calibrated.
■ Meet your application needs with an unsurpassed offering.
Proven best-in-class performance, reliability, and safety
■ More than seven million installed
■ Reference accuracy 0.04 percent of span
■ Installed total performance of 0.14 percent of span
■ 10-year stability of 0.2 percent of URL
■ SIL2/3 certified (IEC 61508)
Feature
- Rosemount Coplanar™plat formenables integration of primary elements,manifolds,and remoteseal solutions
- Best in Class performance with up to 0.05% high accuracy option
- IEC 62591(WirelessHART®)Protocol enables cost effective installations
- Local Operator Interface (LOI) offers easy to use configuration capabilities at the transmitter
- Protocols available include HART®4–20mA,FOUNDATION™Fieldbus,PROFIBUS®PA,HART1–5VdcLow Power
- Selectable HART Revision prepares your plant for the latest HART capabilities while ensuring seamless integration with today's systems
- SIL2/3safety certification to IEC61508 is available with the full 4–20mA HART offering to simplify compliance
Another Article: |
This product data sheet covers HART, Wireless, FOUNDATION Fieldbus, and PROFIBUS PA Protocolsunless specified.
Conformance to specification (±3s[sigma])
Technology leadership, advanced manufacturingtechniques, and statistical process control ensure specification conformance toat least ±3s.
Reference accuracy
Stated reference accuracy equations include terminal based linearity, hysteresis, and repeatability. For Wireless,FOUNDATION Fieldbus, and PROFIBUS PA devices, use calibrated range in place of span.
Models | Standard | High performance option, P8 | |
Rosemount 2051C | |||
Range 1 | ±0.10% of span For spans less than 15:1, accuracy = ±æ0.025+0.005 --U----R---L---ö %ofspan è Spanø |
N/A |
N/A |
Ranges 2–4 |
±0.065% of span For spans less than 10:1, accuracy = ±æ0.025+0.005 --U----R---L---ö %ofspan è Spanø |
Ranges 2–4 | High accuracy option, P8 ±0.05% of span For spans less than 10:1(1), accuracy = ±æ0.015+0.005 --U-----R---L--- ö %ofspan è Spanø |
Range 5 |
±0.075% of span For spans less than 10:1, accuracy= ±æ0.025+0.005 --U-----R---L--- ö %ofspan è Spanø |
Range 5 | High performance option, P8 ±0.065% of span For spans less than 10:1, accuracy= ±æ0.015+0.005 --U----R---L---- ö %ofspan è Spanø |
Rosemount 2051T, 2051G | |||
Ranges 1–4 |
±0.065% of span For spans less than 10:1, accuracy = ±æ0.0075 --U----R---L---ö %ofspan è Spanø |
Ranges 1–4 | High accuracy option, P8 ±0.05% of span For spans less than 10:1(1), accuracy = ±æ0.0075 --U----R---L---- ö %ofspan è Spanø |
Range 5(2) | ±0.075% of span For spans less than 10:1, accuracy = ±æ0.0075 --U----R---L---ö %ofspan è Spanø |
N/A |
N/A |
Rosemount 2051L | |||
Ranges 2–4 | ±0.075% of span For spans less than 10:1, accuracy = ±æ0.025+0.005 --U-----R---L--- ö %ofspan è Spanø |
N/A |
N/A |
1. For protocol code F, accuracy specification is for spans less than 7:1. Not available with out put codeW.
2. Rosemount 2051G is not available with Range5.